ELECTROMED INTERNATIONAL

 

 

       Technical Specifications

 

 

CPX 160

CPX 360

 

 

 

kW

60 kW                                    

  60 kW

kV and mA

 

800 mA at 75 kV

600 mA at 100 kV

400 mA at 125 kV

800 mA at 75 kV

600 mA at 100 kV

400 mA at 125 kV

Radiographic Parameters

 

 

·          kV

40 to 150

40 to 150

·          mA

10 to 800

10 to 800

·          mAs

0.4 to 1000

0.4 to 1000

·          Exposure time

mS to 8 Sec.

1 mS to 8 Sec.

·          Exposure Rate

Up to 16 Exposures per second

Up to 16 Exposures per second

·          Filament Range

Up to 7.0 Amps.

Up to 7.0 Amps.

·          Anode Rotation

Low and/or High Speed

Low and/or High Speed

Fluoroscopic Parameters

 

 

·          kV

40 to 125

40 to 125

·          mA

0.5 to 10 ma in Continuous Fluoro 10 to 50 mA in Pulsed Fluoro

0.5 to 10 ma in Continuous Fluoro 10 to 50 mA in Pulsed Fluoro

·          Pulsed Width

2.5 to 10 mSec.

2.5 to 10 mSec.

·          Timer

5 minute Alarm

5 minute Alarm

Heat Dissipation

600 BTU/Hr.                                         

600 BTU/Hr.

Power Requirements

 

240/480 VAS 60 Hz Single Phase 208 VAC 60 Hz Single Phase       

480 VAC  50/60 Hz Three Phase

Operating Conditions

15° to 40° C at less then 95% humidity

15° to 40° C at less then 95% humidity

X-Ray tube Compatibility

Varian Eimac – Eureka – Comet – Dunlee (other manufacturers upon request)

Varian Eimac – Eureka – Comet – Dunlee (other manufacturers upon request)

Certifications

UL 187 CSA 22.2 N° 114 Compliance to 21 CFR Subchapter J

UL 187; UL 2601; IEC 601-2-7 CSA 22.2 N° 114;  N° 601.1; N° 601.2.7 Compliance to 21 CFR Subchapter J CE Mark

Specifications subject to change without notice

  Home:

 

View Generator Configuration: