ELECTROMED INTERNATIONAL

Technical Specifications
|
|
CPX 160 |
CPX 360 |
|
|
|
|
|
kW |
60 kW |
60 kW |
|
kV and mA
|
800 mA at 75 kV 600 mA at 100 kV 400 mA at 125 kV |
800 mA at 75 kV 600 mA at 100 kV 400 mA at 125 kV |
|
Radiographic Parameters |
|
|
|
· kV |
40 to 150 |
40 to 150 |
|
· mA |
10 to 800 |
10 to 800 |
|
· mAs |
0.4 to 1000 |
0.4 to 1000 |
|
· Exposure time |
mS to 8 Sec. |
1 mS to 8 Sec. |
|
· Exposure Rate |
Up to 16 Exposures per second |
Up to 16 Exposures per second |
|
· Filament Range |
Up to 7.0 Amps. |
Up to 7.0 Amps. |
|
· Anode Rotation |
Low and/or High Speed |
Low and/or High Speed |
Fluoroscopic Parameters |
|
|
|
· kV |
40 to 125 |
40 to 125 |
|
· mA |
0.5 to 10 ma in Continuous Fluoro 10 to 50 mA in Pulsed Fluoro |
0.5 to 10 ma in Continuous Fluoro 10 to 50 mA in Pulsed Fluoro |
|
· Pulsed Width |
2.5 to 10 mSec. |
2.5 to 10 mSec. |
|
· Timer |
5 minute Alarm |
5 minute Alarm |
Heat Dissipation |
600 BTU/Hr. |
600 BTU/Hr. |
|
Power Requirements
|
240/480 VAS 60 Hz Single Phase 208 VAC 60 Hz Single Phase |
480 VAC 50/60 Hz Three Phase |
|
Operating Conditions |
15° to 40° C at less then 95% humidity |
15° to 40° C at less then 95% humidity |
|
X-Ray tube Compatibility |
Varian Eimac – Eureka – Comet – Dunlee (other manufacturers upon request) |
Varian Eimac – Eureka – Comet – Dunlee (other manufacturers upon request) |
Certifications |
UL 187 CSA 22.2 N° 114 Compliance to 21 CFR Subchapter J |
UL 187; UL 2601; IEC 601-2-7 CSA 22.2 N° 114; N° 601.1; N° 601.2.7 Compliance to 21 CFR Subchapter J CE Mark |
Specifications subject to change without notice