International Medical Systems Ziehm |
Siemens Medical Systems |
||||
Model |
Exposcope 7000 |
Exposcope 7000 Compact |
Powermobil |
Siremobil Compact Sub |
Siremobil Iso-C |
|
|
X-ray tube Anode |
Stationary |
Stationary |
Rotating |
Stationary |
Stationary |
|
|
Maximum output |
|
|
|
|
|
|
@120 VAC |
20 mA @ 110 kVp |
20 mA @ 110 kVp |
15 KW ref. to DIN (20 KW) |
1.4 KW |
1.4 KW |
|
@220 VAC |
20 mA @ 110 kVp |
20 mA @ 110 kVp |
15 KW ref. to DIN (20 KW) |
1.4 KW |
1.4 KW |
|
|
Heat capacity, HU |
76,000 |
76,000 |
270,000 |
50,000 |
50,000
|
|
|
Cooling, HU/min |
19,800 |
19,800 |
|
13,700 |
13,700 |
|
|
Focal Spot size, min |
|
|
|
|
|
|
Radiographic Mode |
1.5 |
1.5 |
0.5 |
0.6 |
0.6 |
|
Fluorscopic Mode |
.6 |
.6 |
0.3/0.5 |
0.6 |
0.6 |
|
|
X-Ray Generator |
High Freq., 20 kHz |
High Freq., 20 kHz |
|
|
|
|
|
Power output, KW |
2.2 KW |
2.2 KW |
20 KW |
1.4 KW |
1.4 KW |
|
|
Radiographic mode |
|
|
NA |
|
|
|
kV range |
40-110 |
40-110 |
|
|
|
|
mA range |
20 |
20 |
|
6, 9, 20 mA (Max. 1430 W) |
6, 9, 20 mA |
|
mAs range |
2-80 |
2-80 |
|
1-150 mAs in 23 steps |
1-150 mAs in 23 steps |
|
AEC |
no |
no |
|
mAs |
mAs |
|
|
Exposure time, sec. |
.1 to 4 Sec., 0.1 Increments |
.1 to 4 Sec., 0.1 Increments |
|
0.05-10.0 |
0.05-10.0 |
|
|
Fluoroscopic Mode |
|
|
|
|
|
|
kV range |
40-110 kVp |
40-110 kVp |
40-125 kVp |
40-110 kVp |
40-110 kVp |
|
mA range |
.02 to 8 |
.02 to 8 |
1.8-8.9 |
0.2-8.9 (Max. 550 W) |
0.2-8.9 |
|
Pulsed Fluoro |
yes |
yes |
Yes |
Yes |
Yes |
|
Pulses per sec. |
3 |
3 |
15 |
2 |
2 |
|
ABS Control |
kV, mA, Gain, Black level |
kV, mA, Gain, Black level |
Exposure, TV Iris, Camera gain |
Exposure, TV Iris, Camera gain |
Exposure, TV Iris, Camera gain |
|
Snapshot Mode |
Yes |
Yes |
0.2-250 mA |
0.2-12.2 mA |
0.2-12.2 mA |
|
|
Image Intensifier |
|
|
|
|
|
|
Diameter, inches |
6; 9/6; 12/9/6 Tri-mode |
6; 9/6; 12/9/6 Tri-mode |
9/6 |
7/9 |
7/9 |
|
TV monitor size, in. |
17 |
17 |
17 |
17 |
17 |
|
Cassette holder sizes, Inches. |
10x12 |
10x12 |
|
Cassette Holder Fits 10 x 12 |
Cassette Holder Fits 10 x 12 |
|
|
Image Proc./Storage |
|
|
|
|
|
|
|
Video Storage type |
Digital disc, Digital Memory, Floppy disc, VCR, analog VHS VCR |
Digital Memory |
Digital Memory |
Digital Memory, |
Digital Memory, |
|
Capacity, # of images |
2-16; 32; 352; 852-6952 |
2 |
10,000 |
900 |
900 |
|
Image Matrix size |
1287-512 x 10 bit, 1024 gray scale |
525 x 525 x 8 bit |
512x444 |
512x444 |
512x444 |
|
Last Image Hold |
Yes |
Yes |
Yes |
Yes |
Yes |
|
Frame Integration |
Yes |
Yes |
Yes 2-32 |
Yes 2-32 |
Yes 2-32 |
|
|
Hard Copy Devices |
Multi format camera, single format camera, dye sublimation printer, thermal printer, video printer |
Video Printer |
Paper, Film Printer Multi format camera, Laser camera, DICOM, Removable 1000 Image media |
Paper, Film Printer Multi format camera, Laser camera, DICOM, Removable 1000 Image media |
Paper, Film Printer Multi format camera, Laser camera, DICOM, Removable 1000 Image media |
|
C-Arm |
|
|
|
|
|
|
Free Space, Inches |
30 |
30 |
|
|
|
|
Depth, Inches |
27 |
27 |
|
|
|
|
Orbital rotation, Deg. |
135 |
135 |
190 |
125 |
190 |
|
Horiz. travel, Inches |
9 |
9 |
|
|
|
|
Vert. travel, Inches |
18 |
18 |
|
|
|
|
Panning Motion, Deg. |
+/- 12 |
+/- 12 |
|
|
|
|
Pivot Rotation, Deg. |
+/- 225 |
+/- 225 |
|
|
|
|
Reverse Position |
Yes |
Yes |
|
|
|
|
|
Power Requirements |
115, 20A; 220-240, 16A |
115, 20A; 220-240, 16A |
|
|
|
|
|
H x W x D of Frame, Inches. |
67 x 32 x 63 |
67 x 32 x 63 |
212 x 80 x 181 cm |
165x 80 x 164.8 cm |
212 x 80 x 181 cm |
|
|
Weight, Pounds |
485 |
485 |
|
240 kg. (7”); 250 kg (9”) |
345 kg |
|
|
Warranty |
1 Year |
1 Year |
1 Year |
1 Year |
1 Year |
|
|
Special Features |
Laser positioning, real time edge enhancement & subtraction, Zoom, patient annotation, road mapping, peak opcification, cine loop |
Laser positioning device, Dose measurement |
Digital cine mode up to 250 mA, Cableless counter-balanced iso-centric C-Arm, DSA and road-map standard |
Digital image rotation, electronic shutters on stored images, Image collage, dose area product display, Auto window imaging, Casbleless, Counterbalanced iso-centric C-Arm, user customizable programs, |
Digital image rotation, electronic shutters on stored images, Image collage, dose area product display, Auto window imaging, customizable user programs, Electromagnetic lock controls |
|